|InstrumentName = Nova 600 NanoLab
|InstrumentName = Nova 600
|HeaderColor = #F5A81C
|HeaderColor = #F5A81C
|ImageOne = KNI-Si-and-Pt-Pillar.jpg
|ImageOne = YVO-Nanobeam-Resonator_Jake-Rochman.jpg
|ImageTwo = Nova-NanoLab-600.jpg
|ImageTwo = Nova-NanoLab-600.jpg
|InstrumentType = [[Equipment_List#Microscopy|Microscopy]]
|InstrumentType = [[Equipment_List#Microscopy|Microscopy]]
|RoomLocation = B233B Steele
|RoomLocation = B233B Steele
|LabPhone = 626-395-1534
|LabPhone = 626-395-1534
|PrimaryStaff = [[Matthew S. Hunt, PhD]]
|PrimaryStaff = [[Annalena Wolff]]
|StaffEmail = matthew.hunt@caltech.edu
|StaffEmail = awolff@caltech.edu
|StaffPhone = 626-395-5994
|StaffPhone = 626-395-5994
|Manufacturer = FEI (now Thermo Fisher)
|Manufacturer = FEI (now Thermo Fisher)
|Model = Nova 600 NanoLab
|Techniques = SEM, Ga-FIB, Omniprobe,<br>Immersion Lens Imaging,<br>GIS, Cross-sectioning,<br>TEM Lamella Sample Prep
|Techniques = SEM, Ga-FIB, Omniprobe,<br>Immersion Lens Imaging,<br>GIS, Cross-sectioning,<br>TEM Lamella Sample Prep
|RequestTraining = matthew.hunt@caltech.edu
|RequestTraining = awolff@caltech.edu
|EmailList = kni-sem-fib
|EmailList = kni-sem-fib
|EmailListName = SEM-FIB
|EmailListName = SEM-FIB
}}
}}
== Description ==
== Description ==
The Nova 600 is a "dual beam" system that combines a field emission gun (FEG) scanning electron microscope (SEM) with a gallium focused ion beam (Ga-FIB). It can be used to capture high quality images (with sub-10 nm resolution) and perform site-specific etching and material deposition (with sub-50 nm resolution). It is also equipped with an Omniprobe nanomanipulator, which can be used to lift out lamella samples prepared for use in a transmission electron microscope (TEM). See a full list of training and educational resources for this instrument below.
The Nova 600 is a "dual-beam" system that combines a field emission gun (FEG) scanning electron microscope (SEM) with a gallium focused ion beam (Ga-FIB). It can be used to capture high-quality images (clearly resolving sub-10 nm features) and perform site-specific etching and material deposition (creating sub-20 nm features). It is also equipped with an Omniprobe nanomanipulator, which can be used to lift out lamella samples that are prepared for use in a transmission electron microscope (TEM). See a full list of training and educational resources for this instrument below.
===== SEM Applications =====
===== SEM Applications =====
* Ultra-High-Resolution Imaging (Immersion Mode aka UHR Mode)
* Ultra-High-Resolution Imaging (Immersion Mode aka UHR Mode)
Line 34: Line 35:


== Resources ==
== Resources ==
===== SOPs & Troubleshooting =====
===== Equipment Status =====
* SEM SOPs ([https://caltech.box.com/s/4yv8f5att77k3zq1rm6p0mqhkd8quslh Short Version] | [https://caltech.box.com/s/xwjdudqdl793gkc5kl71zez5nnc5y2xb Long Version])
* [https://labrunr.caltech.edu/Equipment_2.aspx LabRunr Equipment Status] (Select Nova 600 NanoLab from the dropdown menu)
* Ga-FIB SOPs ([https://caltech.box.com/s/8uch8ygjmesmjtdfy65xth47izib0prg Short Version] | [https://caltech.box.com/s/grg6j3rob7c4ciblom1tdshklq62oxqp Long Version])
 
* TEM Lamella Sample Preparation SOPs ([https://caltech.box.com/s/3l3w507dxwosuya3nbxgk30tdqyp4qy9 Short Version] | [https://caltech.box.com/s/pgioaribs7oj8r7g5nncmjjee2uyez2d Long Version])
===== SOPs & Manuals & SDS =====
* [https://caltech.box.com/s/1nmp75l3166vj9t1vwwpwu2zyfc4j6ol Cutting & Imaging Cross-Sections SOP]
* [https://caltech.box.com/s/mpsxkxmf5y8wjw9daijwbudeoqkkeudu KNI Microscopy Policies]
* [https://caltech.box.com/s/sz9pai0icsntnef6me23veiwtwdui0gm Troubleshooting Guide]
* [https://caltech.box.com/s/3v19boseiba88mcujrljctpafyz8f4sh SOP, Manuals and SDS]  
* [https://caltech.box.com/s/og4309108q4k2jwhkaxqtpiujg2al5iu Nova NanoLab Operation Manual]
* [https://caltech.box.com/s/j0t3w6i53jhfjcva8i4qvlatdh7t1tzw Gas Injection Systems – Deposition of Platinum (Technical Note)]
* [https://caltech.box.com/s/sm7q7teh5fo5hg3e6flkjbcbycmarrm3 Gas Injection Systems – Deposition of SiOx (Technical Note)]
* [https://caltech.box.com/s/4pcym0l9j1e8t9b2vznrzps9uk7b85mh Gas Injection Systems – Etching with IEE aka XeF<sub>2</sub> Etch (Technical Note)]
* [https://caltech.box.com/s/110tb0o8avjziwa1y4d017dbbcpkxfop Scripting – AutoScript Language Manual (year 2000 Technical Note: most complete)]
* [https://caltech.box.com/s/tlqgvtkkiahi261megm087i61gqlfzrc Scripting – RunScript Manual]
 
===== Video Tutorials =====
===== Video Tutorials =====
* [https://youtu.be/UfF_ljwvepQ Getting Started] | [https://youtu.be/luC-5TgNPsQ Basic SEM Alignment]
* [https://youtu.be/UfF_ljwvepQ Getting Started] | [https://youtu.be/luC-5TgNPsQ Basic SEM Alignment]
* Astigmatism Correction ([https://youtu.be/YeukVt1Fyi0 Details] | [https://youtu.be/WFfOi-rwlbA On Right-Angle Features] | [https://youtu.be/1syySgnTEqU Stigmator Alignment])
* Astigmatism Correction ([https://youtu.be/YeukVt1Fyi0 Details] | [https://youtu.be/WFfOi-rwlbA On Right-Angle Features] | [https://youtu.be/1syySgnTEqU Stigmator Alignment])
* [https://youtu.be/4V-bE6uqHY4 Eucentric Height: What it means, When to use it & How to get there]
* [https://youtu.be/R_RYbtumU20 Adjusting TLD Voltage to Capture SE vs. BSE Signal]
* TEM Lamella Sample Prep ([https://www.youtube.com/playlist?list=PL7Lb5X_YIzOkg3wRe6A5a5b76fFxYyT3s Playlist])
 
* Cutting & Imaging Cross-sections ([https://www.youtube.com/playlist?list=PL7Lb5X_YIzOnW6dD0GHeasXS6MRVIgoTA Playlist])
* [https://youtu.be/67RdSeJcaGs Milling Non-Conductive Samples using Charge Compensation]
* [https://youtu.be/XNgYe8ZDJvo Perfecting Ga-FIB Alignments]
* [https://youtu.be/R_RYbtumU20 Adjusting TLD Voltage to Capture SE vs. BSE Signal]  
===== Graphical Handouts =====
===== Graphical Handouts =====
* [https://caltech.box.com/s/14ffgscc39vhrvlyva0jbucsep7j6dvv SEM Concepts]
* [https://caltech.box.com/s/pxl99bbc1jm1tbjshfaotm91xm0mqs1i SEM Concepts]
* [https://caltech.box.com/s/kxaxtslwol1o5a276f3lrqbhss8zvwje Ga-FIB Concepts]
* [https://caltech.box.com/s/kxaxtslwol1o5a276f3lrqbhss8zvwje Ga-FIB Concepts]
* [https://caltech.box.com/s/k2iy75hxiwkehv0ogqelz2sux9e1cs5k SEM & Ga-FIB Alignments]
* [https://caltech.box.com/s/k2iy75hxiwkehv0ogqelz2sux9e1cs5k SEM & Ga-FIB Alignments]
* [https://caltech.box.com/s/ijd8gprg9gcavb6of5uegu7osinzsdet Guide to Optimizing SEM Imaging]
* [https://caltech.box.com/s/ijd8gprg9gcavb6of5uegu7osinzsdet Guide to Optimizing SEM Imaging]
===== Presentations =====
===== Presentations =====
* [https://caltech.box.com/s/lulkj0pwm053akyya1shazg8wzgudq9f Scanning Electron Microscopy: Principles, Techniques & Applications]
* Scanning Electron Microscopy (SEM): Principles, Techniques & Applications
* [https://caltech.box.com/s/f4k8jan85n5lf6f2tutjx4rkfzjq7y68 Gallium Focused Ion Beam Microscopy: Principles, Techniques & Applications]
** [https://caltech.box.com/s/lulkj0pwm053akyya1shazg8wzgudq9f Pptx Slides] | [https://youtu.be/Zh21tp3aPEw YouTube Lecture]
===== Manufacturer Manuals =====
* Gallium Focused Ion Beam (Ga-FIB) Microscopy: Principles, Techniques & Applications
* [https://caltech.box.com/s/og4309108q4k2jwhkaxqtpiujg2al5iu Nova NanoLab Operation Manual]
** [https://caltech.box.com/s/f4k8jan85n5lf6f2tutjx4rkfzjq7y68 PPtx Slides] | [https://youtu.be/3eSzisbNcGo YouTube Lecture]
* [https://caltech.box.com/s/36j0n01ig7f757kvkiebvqjqj0el2yvp EDAX Genesis EDS & WDS Operation Manual]
 
* [https://caltech.box.com/s/j0t3w6i53jhfjcva8i4qvlatdh7t1tzw Gas Injection Systems – Deposition of Platinum (Technical Note)]
 
* [https://caltech.box.com/s/sm7q7teh5fo5hg3e6flkjbcbycmarrm3 Gas Injection Systems – Deposition of SiOx (Technical Note)]
* [https://caltech.box.com/s/4pcym0l9j1e8t9b2vznrzps9uk7b85mh Gas Injection Systems – Etching with IEE aka XeF2 Etch (Technical Note)]
* [https://caltech.box.com/s/b95pg59k003i9vn8y9elzem0tnxqg3fb Gas Injection Systems – Beam Chemistries Presentation]
* [https://caltech.box.com/s/110tb0o8avjziwa1y4d017dbbcpkxfop Scripting – AutoScript Language Manual (year 2000 Technical Note: most complete)]
* [https://caltech.box.com/s/n0abqy5z1e9a8hbyf38qybxu86sqz3vb Scripting – AutoScript Language Manual (year 2005 Technical Note: less complete, still useful)]
* [https://caltech.box.com/s/tlqgvtkkiahi261megm087i61gqlfzrc Scripting – RunScript Manual]
===== Simulation Software =====
===== Simulation Software =====
* [http://www.gel.usherbrooke.ca/casino/What.html CASINO Electron Beam Simulation Software – simulate e-beam/specimen interactions]
* [http://www.gel.usherbrooke.ca/casino/What.html CASINO Electron Beam Simulation Software – simulate e-beam/specimen interactions]
Line 73: Line 72:
* The KNI has a NIST-traceable standard against which SEM and Ga-FIB measurements can be compared. See Slides 54-55 of the [https://caltech.box.com/s/lulkj0pwm053akyya1shazg8wzgudq9f SEM Presentation] for details. Ask staff for help finding and using the standard in the lab.
* The KNI has a NIST-traceable standard against which SEM and Ga-FIB measurements can be compared. See Slides 54-55 of the [https://caltech.box.com/s/lulkj0pwm053akyya1shazg8wzgudq9f SEM Presentation] for details. Ask staff for help finding and using the standard in the lab.
===== Sample Preparation =====
===== Sample Preparation =====
* Use the [[Leica EM ACE600: Carbon Evaporator | Carbon Evaporator]] to make non-conductive samples conductive by applying 2-10 nm of evaporated carbon.
* Use the [[Carbon Evaporator]] to make non-conductive samples conductive by applying 2-10 nm of evaporated carbon.
* Use the [[Tergeo Plus ICP- & CCP-RIE: Oxygen & Argon Plasma Cleaner |
* Use the [[Tergeo Plus ICP- & CCP-RIE: Oxygen & Argon Plasma Cleaner |
O<sub>2</sub>/Ar Plasma Cleaner]] to remove hydrocarbons from the sample surface to avoid creating dark contamination spots on your features while imaging them.
O<sub>2</sub>/Ar Plasma Cleaner]] to remove hydrocarbons from the sample surface to avoid creating dark contamination spots on your features while imaging them.
===== Order Your Own Stubs =====
 
* Stubs used for mounting specimens are considered a personal, consumable item in the KNI. There are some old stubs at each SEM/FIB, yet you should buy your own so that you can keep them clean and available to you. There are many stub geometries and configurations, some of which will be right for you to purchase and keep with your other cleanroom items.
===== Stubs for specimen mounting =====
** [https://www.tedpella.com/SEM_html/SEMclip.htm.aspx Buy stubs with copper clips] (recommended for most devices, esp with non-conductive substrates);
* Stubs used for mounting specimens are considered a personal, consumable item in the KNI. There are some stubs at each Microscope which can be used by any KNI microscopy user. You can also buy your own stubs so that you can keep them clean and available to you. There are many stub geometries and configurations. If you chose to buy your own stubs, please show them to the staff microscopist prior to using them: some stubs including stubs with copper clips have large height differences and can only be used safely in specific operating conditions.
** [https://www.tedpella.com/sem_html/SEMpinmount.htm Buy stubs without copper clips] (ok for devices with conductive substrates)
 
** [https://www.tedpella.com/sem_html/SEMpinmount.htm Buy stubs without copper clips]
 
===== Guide to Choosing KNI SEMs & FIBs =====
*[[Guide to Choosing KNI SEMs & FIBs | Consult this guide for help in choosing the best SEMs and FIBs for your work]]


== Specifications ==
== Specifications ==
Line 90: Line 93:
* Apertures: 10 &mu;m, 15 &mu;m, 20 &mu;m, 30 &mu;m
* Apertures: 10 &mu;m, 15 &mu;m, 20 &mu;m, 30 &mu;m
* Eucentric Height: ~5.15 mm working distance (WD)
* Eucentric Height: ~5.15 mm working distance (WD)
* Stage Range: +/- 80 mm X & Y travel, 12 mm Z travel, -12 to 58&deg; tilt, 360&deg; rotation
* Stage Range: ±80 mm X & Y travel, 12 mm Z travel, -12 to 58&deg; tilt, 360&deg; rotation
* ETD Grid Bias Range: -150 to 300 V
* ETD Grid Bias Range: -150 to 300 V
* TLD Bias Range: -100 to 150 V
* TLD Bias Range: -100 to 150 V
Line 104: Line 107:
* ETD Grid Bias Range: -150 to 300 V
* ETD Grid Bias Range: -150 to 300 V
* TLD Bias Range: -100 to 150 V
* TLD Bias Range: -100 to 150 V
<br>
<br>

Revision as of 05:19, 30 June 2022

Nova 600
YVO-Nanobeam-Resonator Jake-Rochman.jpg
Instrument Type Microscopy
Techniques SEM, Ga-FIB, Omniprobe,
Immersion Lens Imaging,
GIS, Cross-sectioning,
TEM Lamella Sample Prep
Staff Manager Annalena Wolff
Staff Email awolff@caltech.edu
Staff Phone 626-395-5994
Reserve time on LabRunr
Request training by email
Lab Location B233B Steele
Lab Phone 626-395-1534
Manufacturer FEI (now Thermo Fisher)
Model Nova 600 NanoLab
Nova-NanoLab-600.jpg

Description

The Nova 600 is a "dual-beam" system that combines a field emission gun (FEG) scanning electron microscope (SEM) with a gallium focused ion beam (Ga-FIB). It can be used to capture high-quality images (clearly resolving sub-10 nm features) and perform site-specific etching and material deposition (creating sub-20 nm features). It is also equipped with an Omniprobe nanomanipulator, which can be used to lift out lamella samples that are prepared for use in a transmission electron microscope (TEM). See a full list of training and educational resources for this instrument below.

SEM Applications
  • Ultra-High-Resolution Imaging (Immersion Mode aka UHR Mode)
  • High-Resolution Imaging (Field-Free Mode aka Normal Mode)
  • Secondary Electron (SE) imaging with an Everhart-Thornley Detector (ETD) & Through-the-Lens Detector (TLD)
  • Backscattered Electron (BSE) imaging with a TLD
  • Platinum deposition via Gas Injection System (GIS)
  • Automated imaging with RunScript program & AutoScript language
Ga-FIB Applications
  • Directly etch patterns into material
  • Cutting & Imaging Cross-Sections
  • TEM Lamella Sample Preparation using an Omniprobe for Liftout
  • Platinum & SiOx deposition via GIS
  • Enhanced etch with XeF2 via GIS
  • Automated patterning with RunScript program & AutoScript language

Resources

Equipment Status
SOPs & Manuals & SDS
Video Tutorials
Graphical Handouts
Presentations


Simulation Software
Calibrate Measurements with NIST Standard
  • The KNI has a NIST-traceable standard against which SEM and Ga-FIB measurements can be compared. See Slides 54-55 of the SEM Presentation for details. Ask staff for help finding and using the standard in the lab.
Sample Preparation
  • Use the Carbon Evaporator to make non-conductive samples conductive by applying 2-10 nm of evaporated carbon.
  • Use the O2/Ar Plasma Cleaner to remove hydrocarbons from the sample surface to avoid creating dark contamination spots on your features while imaging them.
Stubs for specimen mounting
  • Stubs used for mounting specimens are considered a personal, consumable item in the KNI. There are some stubs at each Microscope which can be used by any KNI microscopy user. You can also buy your own stubs so that you can keep them clean and available to you. There are many stub geometries and configurations. If you chose to buy your own stubs, please show them to the staff microscopist prior to using them: some stubs including stubs with copper clips have large height differences and can only be used safely in specific operating conditions.
Guide to Choosing KNI SEMs & FIBs

Specifications

Manufacturer Specifications
SEM Specifications
  • Minimum Feature Size Resolved in Immersion Mode: ~5 nm
  • Voltage Range: 0.2 to 30.0 kV
  • Current Range: ~10 pA to 20 nA
  • Apertures: 10 μm, 15 μm, 20 μm, 30 μm
  • Eucentric Height: ~5.15 mm working distance (WD)
  • Stage Range: ±80 mm X & Y travel, 12 mm Z travel, -12 to 58° tilt, 360° rotation
  • ETD Grid Bias Range: -150 to 300 V
  • TLD Bias Range: -100 to 150 V
  • Ultimate Vacuum: 5e-7 mbar
Ga-FIB Specifications
  • Minumum Probe Size Achieved: ~7 nm
  • Minimum Feature Size Etched: ~25 nm
  • Minimum Feature Size Resolved by Imaging: ~10 nm
  • Voltage Range: 5 to 30 kV
  • Current Range: 1 pA to 20 nA
  • Eucentric Height: ~5.15 mm working distance (WD)
  • Stage Tilt to be perpendicular to Ga-FIB: 52°
  • ETD Grid Bias Range: -150 to 300 V
  • TLD Bias Range: -100 to 150 V