|RoomLocation = B242A Keck | |RoomLocation = B242A Keck | ||
|LabPhone = 626-395-8908 | |LabPhone = 626-395-8908 | ||
|PrimaryStaff = [[ | |PrimaryStaff = [[Guy A. DeRose, PhD]] | ||
|StaffEmail = | |StaffEmail = derose@caltech.edu | ||
|StaffPhone = 626-395- | |StaffPhone = 626-395-3423 | ||
|Manufacturer = Assorted (see list on page) | |Manufacturer = Assorted (see list on page) | ||
|Model = Assorted (see list on page) | |Model = Assorted (see list on page) | ||
|Techniques = Cross-Section Sample Prep,<br>Sample Polishing,<br>3 mm Core Drilling,<br>Dimpling, Argon Milling | |Techniques = Cross-Section Sample Prep,<br>Sample Polishing,<br>3 mm Core Drilling,<br>Dimpling, Argon Milling | ||
|RequestTraining = | |RequestTraining = derose@caltech.edu | ||
|EmailList = kni-tf30 | |EmailList = kni-tf30 | ||
|EmailListName = TF-30 | |EmailListName = TF-30 | ||
Line 52: | Line 52: | ||
===== Transmission Electron Microscopes ===== | ===== Transmission Electron Microscopes ===== | ||
* [[Tecnai TF-30: 300 kV TEM, STEM, EDS & HAADF | Tecnai TF-30: TEM, STEM, EDS & HAADF (50-300 kV)]] | * [[Tecnai TF-30: 300 kV TEM, STEM, EDS & HAADF | Tecnai TF-30: TEM, STEM, EDS & HAADF (50-300 kV)]] | ||
<!--- | |||
* [[Tecnai TF-20: 200 kV TEM, STEM, EDS, EELS, EFTEM & Lithography | Tecnai TF-20: TEM, STEM, EDS, EELS, EFTEM & Lithography (40-200 kV)]] | * [[Tecnai TF-20: 200 kV TEM, STEM, EDS, EELS, EFTEM & Lithography | Tecnai TF-20: TEM, STEM, EDS, EELS, EFTEM & Lithography (40-200 kV)]] | ||
---> | |||
===== Sample Preparation for TEM ===== | ===== Sample Preparation for TEM ===== | ||
* [[Nova 600 NanoLab: SEM, Ga-FIB, GIS & Omniprobe|Nova 600 NanoLab: SEM, Ga-FIB, GIS & Omniprobe]] | * [[Nova 600 NanoLab: SEM, Ga-FIB, GIS & Omniprobe|Nova 600 NanoLab: SEM, Ga-FIB, GIS & Omniprobe]] |
Revision as of 20:52, 8 February 2022
|
Description
The KNI offers a lab in the Keck sub-basement for preparing TEM samples by traditional means, often to produce cross-section specimens from a bulk sample. This typically involves the following steps:
- Glue together a stack of materials, e.g. with the surface of two samples facing each other
(if the surface is what you want to observe in the TEM) - Cut the stack into thin slices of material, in the appropriate orientation
(the thin dimension here will be the same thin dimension in the TEM) - Cut out a 3 mm circle from this thin slice using a Disk Cutter
- Polish that thin disk down to a thickness of ~100 μm
- Use a Dimpler to polish a hemispherical dimple into the middle of the disk,
such that only a few microns of material remain at the thinnest portion - Use an Argon Mill to perform the final polish of the thinnest portion,
until a hole has been bored through the middle and
the edges of that hole are thin enough to be electron-transparent in the TEM
List of Equipment Available
- Low-Angle Argon Milling & Polishing Station: Fischione Model 1010 (shown in infobox image)
- Dimpler: Fischione Model 2000
- Ultrasonic Disk Cutter: Model 170
- Polishing Station: Allied TechPrep
- Polishing Station: Buehler Ecomet 3
- Low Speed Saw: Isomet
- Roll Grinder: Buehler Handimet 2
- Various polishing media & accessories:
- Media: grit paper, lapping films, polishing cloths, diamond suspensions, alumina poweders
- Accessories: tripod polisher (South Bay Technology), endpoint polisher (Fischione Model 160)
Resources
SOPs
- Paper copies of SOPs are available in the lab for the Argon Ion Mill & Dimpler